6 Kmsrup A/S • 5512850_D1_GB_01.2016
MULTICAL® 402
1 General information
Read these instructions before you install the meter.
In case of incorrect mounting Kamstrup’s guarantee obligations no longer
apply.
Plese noe h he following insllion condiions mus be obeyed:
- Pressure sge: PN16/PN25, see mrking. The flow
sensor mrking does no pply o
enclosed ccessories.
- Pressure sge, Kmsrup
sensor se ype DS: PN16
- Pressure sge, Kmsrup
sinless seel sensor pockes: PN25
A  medium emperure bove 90 °C we recommend he use of flnge
sensors s well s wll mouning of he clculor.
2 Mounting of temperature sensors
The emperure sensors used for mesuring forwrd nd reurn
emperures respecively, consiue  mched sensor pir, which mus
never be sepred. Temperure sensors re usully mouned in he
clculor from he fcory. According o EN 1434/OIML R75 he cble lengh
mus no be chnged. Should replcemen be necessry, boh sensors mus
be replced.
The sensor mrked wih  red sign is o be inslled in he flow pipe. The
oher sensor, mrked wih  blue sign, is o be inslled in he reurn pipe.
For mouning in he clculor, see he prgrph ”Elecricl connecion.
Note: The sensor cbles mus neiher be exposed o jerking nor pulling.
Plese be wre of his when binding he cbles, nd be creful no o
pull he binders unnecessrily igh s his my dmge he cbles.
Plese lso noe h emperure sensors mus be mouned from
below in cooling or he/cooling insllions.