10 Kmsrup A/S • 5512850_D1_GB_01.2016
MULTICAL® 402
4 Mounting of flow sensor
Prior o insllion of he flow sensor, he sysem should be flushed nd
proecion plugs/plsic diphrgms removed from he flow sensor.
Correc flow sensor posiion (inle or oule) ppers from he fron lbel of
MULTICAL® 402. The flow direcion is indiced by n rrow on he side of he
flow sensor.
4.1 Mounting of couplings as well as short direct sensor mounted in
MULTICAL®402 flow sensor
The shor direc sensor from
Kmsrup my be mouned in
PN16 insllions only, wheres
he blind plug which is mouned
in he MULTICAL® 402 flow sensor
from he fcory my be used in
connecion wih boh PN16 nd
PN25 insllions.
The flow sensor my be used in
boh PN16 nd PN25 insllions
nd is vilble mrked eiher PN16
or PN25 s required.
Enclosed couplings, if ny, re only
inended for PN16. Suible PN25
couplings mus be used in PN25
insllions.
In connection with G3/4x110 mm
and G1x110 mm it must be checked
that 10 mm thread run-out is
sufficient. See the figure to the
right.
Gske
Momen pprox. 4 Nm
Gske
10 mm