EasyManuals Logo

KaVo PROPHYflex 4 User Manual

KaVo PROPHYflex 4
34 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #14 background imageLoading...
Page #14 background image
Instructions for use PROPHYflex 4
4 Start up and shut down | 4.1 Installing the MULTIflex coupling
14 / 34
4 Start up and shut down
WARNING
Hazard from non-sterile products.
Infection hazard for dentist and patient.
Before first use and after each use, prepare and sterilise the medical device
and accessories appropriately.
WARNING
Dispose of the product in appropriate manner.
Infection hazard.
Before disposal, reprocess and sterilise the product and accessories appro-
priately.
NOTICE
Damage from soiled and moist cooling air.
Contaminated and moist cooling air can cause malfunctions.
Make sure that the supply of cooling air is dry, clean, and uncontaminated
according to EN ISO 7494-2.
4.1 Installing the MULTIflex coupling
WARNING
Detachment of the medical device during treatment.
A medical device that is not properly locked can release from the MULTIflex
coupling during treatment.
Before each use, check if the medical device is securely locked onto the
MULTIflex coupling.
Screw the MULTIflex coupling onto the turbine hose.
Open the water supply all the way using the spray ring on the MULTIflex
coupling.
4.2 Check the O-rings (MULTIflex)
NOTICE
Missing or damaged O-rings.
Malfunction and premature failure.
Make sure that all O-rings are on the coupling and are undamaged.
Number of available O-rings: 5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KaVo PROPHYflex 4 and is the answer not in the manual?

KaVo PROPHYflex 4 Specifications

General IconGeneral
BrandKaVo
ModelPROPHYflex 4
CategoryMedical Equipment
LanguageEnglish

Related product manuals