EasyManua.ls Logo

KaVo PROPHYflex 4 - Steam Sterilization (Autoclaving)

KaVo PROPHYflex 4
34 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instructions for use PROPHYflex 4
7 Reprocessing steps in accordance with ISO 17664 | 7.7 Sterilisation
28 / 34
7.7 Sterilisation
Sterilisation in a steam steriliser (autoclave) in
accordance with EN 13060 / ISO 17665-1
NOTICE
Contact corrosion due to moisture.
Damage to product.
Immediately remove the product from the steam steriliser after the steril-
isation cycle.
Note
Prior to attaching the powder container, all powder-conducting parts and air
channels must be absolutely dry. Screw together the powder container and
handpiece only in the cold state.
The KaVo medical device has a maximum temperature resistance up to 138
(280.4 °F).
Select a suitable procedure (depending on the available autoclave) from the fol-
lowing sterilisation processes:
Steriliser with triple pre-vacuum:
- at least 3 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4 °F)
Steriliser using the gravity method:
- at least 10 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4 °F) or
alternatively
- at least 60 minutes at 121 °C -1 °C / +4 °C (250 °F -1.6 °F / +7.4 °F)
Use according to the manufacturer's Instructions for Use.
7.8 Storage
Reprocessed products must be stored appropriately such that they are protec-
ted from germs (as far as possible) and dust, in a dry, dark, cool room.
Note
Comply with the expiry date of the sterilised items.

Related product manuals