6 Troubleshooting | 6.1 Cleaning a blocked cannula
6 Troubleshooting
Preventive measures
After each treatment and before each sterilisation, unscrew the powder
container in an anticlockwise direction and replace it with a clean powder
container.
Mount the PROPHYflex on the MULTIflex coupling and blow through the air
and water channels.
Switch the water off and blow through the air and water channels again.
6.1 Cleaning a blocked cannula
Unscrew the cannula in anticlockwise direction using the wrench.
Slide the nozzle needle into the cannula from the front while rotating it.
Then slide the nozzle needle into the cannula from the back while rotating
it.
Then remove the nozzle needle, and blow out the cannula with compressed
air.