Instructions for use PROPHYflex 4
7 Reprocessing steps in accordance with ISO 17664 | 7.4 Automated reprocessing
26 / 34
7.4.2 Pre-cleaning
Accessories required:
▪ Tap water 30
o
C ± 2
o
C (86
o
F ± 4
o
F)
▪ Brush, e.g. medium-hard toothbrush
▶ Disassemble the instrument completely.
▶ Brush off all individual parts under running tap water.
7.4.3 Preparation for automated internal and external
cleaning as well as internal and external disinfection
Miele Series G 7881/7891
Main body required material:
Cleaning cover PROPHYflex 4
(3.004.6658)
Reprocessing with Miele AUF Ad-
apter
Gripping sleeve required material:
Cleaning in Miele sieve basket
Cannula / Adapter + Perio Tip required material:
Cleaning adapter PROPHYflex 4
long (3.004.6640)
Reprocessing with Miele AUF Ad-
apter
Gripping sleeve and cannula /
Adapter + Perio Tip
Reprocessing with Miele AUF Ad-
apter
Powder container required material:
Reprocessing in Miele sieve bas-
ket