Instructions for use PROPHYflex 4
7 Reprocessing steps in accordance with ISO 17664 | 7.4 Automated reprocessing
26 / 34
7.4.2 Pre-cleaning
Accessories required:
Tap water 30
o
C ± 2
o
C (86
o
F ± 4
o
F)
Brush, e.g. medium-hard toothbrush
Disassemble the instrument completely.
Brush off all individual parts under running tap water.
7.4.3 Preparation for automated internal and external
cleaning as well as internal and external disinfection
Miele Series G 7881/7891
Main body required material:
Cleaning cover PROPHYflex 4
(3.004.6658)
Reprocessing with Miele AUF Ad-
apter
Gripping sleeve required material:
Cleaning in Miele sieve basket
Cannula / Adapter + Perio Tip required material:
Cleaning adapter PROPHYflex 4
long (3.004.6640)
Reprocessing with Miele AUF Ad-
apter
Gripping sleeve and cannula /
Adapter + Perio Tip
Reprocessing with Miele AUF Ad-
apter
Powder container required material:
Reprocessing in Miele sieve bas-
ket