Instructions for use PROPHYflex 4
6 Troubleshooting | 6.2 Cleaning a clogged main body
20 / 34
6.2 Cleaning a clogged main body
Pulling off the grip sleeve with cannula.
Unscrew the powder container anticlockwise.
Push the nozzle needle through the aperture of the nozzle.
Use the cleaning drill to clean or remove obstructions from the media tube.
Then blow through with compressed air.
If the clogging persists:
Use the wrench for cannula to unscrew the nozzle by placing the lateral re-
cess of the wrench on the nozzle and unscrewing the nozzle.