Instructions for use PROPHYflex 4
7 Reprocessing steps in accordance with ISO 17664 | 7.4 Automated reprocessing
24 / 34
7.4 Automated reprocessing
WARNING
Incomplete disinfection.
Infection hazard.
Only use disinfection procedures that are verified to be bactericidal, fungi-
cidal and virucidal.
If the disinfectants used do not meet these requirements, the process must
be concluded by disinfection of the unit(s) without packaging using a steam
steriliser.
NOTICE
Never disinfect the handpiece with chloride-containing products.
Malfunction and material damage.
Only disinfect in a washer disinfector or manually.
NOTICE
Never reprocess this medical device in an ultrasonic device.
Malfunction and material damage.
Clean manually or in a washer disinfector only.