Chapter 8. Starting Standalone Manual Calibration
______________________________________________________________________________
CAUTION! Risk of burns!
The calibrator can become very hot when in operation. Touching hot parts can result in
serious injuries.
Never touch the metal block, the tank, the adapter sleeve or the test specimen at
temperatures above 35 °C or below 10 °C.
Allow the calibrator to cool before you remove the test specimen, clean the tank, change
the adapter sleeve or the calibration insert or switch off the machine.
______________________________________________________________________________
Before the testing process:
Before the start of the testing process, check whether
 the instructions regarding the installation site and the operational position have been complied
with
 the electrical connections have been made correctly.
 the correct measuring insert has been selected for the test task.
 the test sensors are securely fixed in the calibrator.
 the calibrator has sufficient structural stability.
Starting the testing process:
IMPORTANT! Parameter fields disabled!
During the testing process, all the parameter fields, except for the Start / Stop slider and the
alarm signal, are disabled.
 Push the button of the start / stop slider completely to the right.
* The testing process is started. Information on the testing
process and the temperature progression is displayed in the window.
Chapter 8. Starting Standalone Manual Calibration