GETTING STARTED
17
Rising current slew rate F-06
0.01A/s~72.00A/s (2260B-30-36)
0.1A/s~144.0A/s (2260B-30-72)
0.01A/s~27.00A/s (2260B-80-13)
0.01A/s~54.00A/s (2260B-80-27)
Falling current slew rate F-07
0.01A/s~72.00A/s (2260B-30-36)
0.1A/s~144.0A/s (2260B-30-72)
0.01A/s~27.00A/s (2260B-80-13)
0.01A/s~54.00A/s (2260B-80-27)
Internal resistance
setting
F-08
0.000Ω~0.833Ω (2260B-30-36)
0.000Ω~0.417Ω (2260B-30-72)
0.000Ω~5.926Ω (2260B-80-13)
0.000Ω~2.963Ω (2260B-80-27)
Bleeder circuit control F-09 0 = OFF, 1 = ON
Buzzer ON/OFF control
F-10 0 = ON, 1 = OFF
USB/GPIB settings
Front panel USB State F-20 0 = Absent, 1 = Mass Storage
Rear panel USB State F-21
0 = Absent, 2 = USB-CDC, 3 = GPIB-
USB adapter
Rear panel USB mode F-22
0 = Disable, 1 = GPIB-USB adapter,
2 = USB CDC
GPIB address F-23 0~30
LAN settings
MAC Address-1 F-30 0x00~0xFF
MAC Address-2 F-31 0x00~0xFF
MAC Address-3 F-32 0x00~0xFF
MAC Address-4 F-33 0x00~0xFF
MAC Address-5 F-34 0x00~0xFF
MAC Address-6 F-35 0x00~0xFF
LAN F-36 0 = Disable, 1 = Enable
DHCP F-37 0 = Disable, 1 = Enable
IP Address-1 F-39 0~255
IP Address-2 F-40 0~255
IP Address-3 F-41 0~255
IP Address-4 F-42 0~255
Subnet Mask-1 F-43 0~255
Subnet Mask-2 F-44 0~255
Subnet Mask-3 F-45 0~255
Subnet Mask-4 F-46 0~255
Gateway-1 F-47 0~255
Gateway-2 F-48 0~255