2260B Series Programming Manual
66
Return parameter
<Boolean>
Returns power switch trip setting.
SYSTem:CONFigure:CURRent:CONTrol
Description
Sets or queries the CC control mode (local control
(panel), external voltage control, external
resistance control). This setting is applied only
after the power is cycled.
Syntax
Query Syntax
SYSTem:CONFigure:CURRent:CONTrol { 0 | 1 | 2 | 3 }
SYSTem:CONFigure:CURRent:CONTrol?
Parameter/Return <NR1>
Description
0
Local (Panel) control
1
External voltage control
2
External resistance control; 10k = Io max,
0k = Io min.
3
External resistance control; 10k = Io min,
0k = Io max.
SYSTem:CONFigure:VOLTage:CONTrol
Description
Sets or queries the CV control mode (local control,
external voltage control, external resistance
control). This setting is applied only after the
power is cycled.
Syntax
Query Syntax
SYSTem:CONFigure:VOLTage:CONTrol { 0 | 1 | 2 | 3 }
SYSTem:CONFigure:VOLTage:CONTrol?
Parameter/Return <NR1>
Description
0
Local (Panel) control
1
External voltage control
2
External resistance control; 10k = Vo max,
0k = Vo min.
3
External resistance control; 10k = Vo min,
0k = Vo max.