REMOTE CONTROL
55
Source Commands
[SOURce:]CURRent[:LEVel
]
[:IMMediate][:AMPLitude]55
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude] . 56
[SOURce:]CURRent:PROTection[:LEVel] ...................... 56
[SOURce:]CURRent:PROTection:STATe ........................ 57
[SOURce:]CURRent:SLEW:RISing ................................... 57
[SOURce:]CURRent:SLEW:FALLing ............................... 57
[SOURce:]RESistance[:LEVel][:IMMediate][:AMPLitude]
................................................................................................. 58
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]
................................................................................................. 58
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude] 59
[SOURce:]VOLTage:PROTection[:LEVel] ...................... 59
[SOURce:]VOLTage:SLEW:RISing ................................... 60
[SOURce:]VOLTage:SLEW:FALLing .............................. 60
[SOURce:]CURRent[:LEVel][:IMMediate]
[:AMPLitude]
Description
Sets or queries the current level in amps. For
externally set current levels (from the analog
control connector) the set current level is returned.
Syntax [SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]
{<NRf>|MIN|MAX}
Query Syntax [SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]?
[MIN|MAX]
Parameter/Return
<NRf>
0~105% of the rated current output level.
MIN
Minimum current level.
MAX
Maximum current level.
Example SOUR:CURR:LEV:IMM:AMPL? MAX
37.800
Returns the maximum possible current level in amps.
Set
Query

Table of Contents