EasyManuals Logo

KERN 440 Series User Manual

KERN 440 Series
Go to English
110 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #74 background imageLoading...
Page #74 background image
440-BA-e-1744
36
12 Instant help
In case of an error in the program process, briefly turn off the balance and disconnect
from power supply. The weighing process must then be restarted from the beginning.
Help:
Fault Possible cause
The displayed weight does not glow.
The balance is not switched on.
The mains supply connection has been
interrupted (mains cable not plugged
in/faulty).
Power supply interrupted.
Batteries are inserted incorrectly or empty
No batteries inserted.
The displayed weight is permanently
changing
Draught/air movement
Table/floor vibrations
The weighing plate is in contact with
foreign matter.
Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
The weighing value is obviously
wrong
The display of the balance is not at zero
Adjustment is no longer correct.
Great fluctuations in temperature.
Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN 440 Series and is the answer not in the manual?

KERN 440 Series Specifications

General IconGeneral
Weighing plate materialstainless steel
Readability0.01g
CalibrationExternal
DisplayLCD
Reproducibility0.01 g - 1 g
Minimum part weight at piece counting (Laboratory scale)0.01g
Minimum part weight at piece counting (Precision scale)0.01g
Power supply/adapter includedincluded
Weighing unitsg, kg, oz, lb
InterfaceRS232
Unitsg, kg, oz, lb
Stabilisation time under laboratory conditions2 seconds

Related product manuals