1 Technical Data
KERN 440-21N 440-21A 440-33N
Readability (d) 0.001 g 0.001 g 0.01 g
Weighing range (max) 40 g 60 g 200 g
Taring range (subtractive) 40 g 60 g 200 g
Reproducibility 0.001 g 0.001 g 0.01 g
Linearity ± 0.003 g ± 0.003 g ± 0.02 g
Minimum unit weight at
piece counting
0.002 g 0.002 g 0.02 g
Warm-up time 2 hours 2 hours 30 minutes
Reference quantities at
piece counting
5, 10, 25, 50
Weighing Units Details „Weighing units“ chapter 9.3
Recommended adjustment weight,
not added (class)
40g (F1) 50g (F2) 200g (M1)
Stabilization time (typical) 3 sec.
Operating temperature + 5° C .... + 35° C
Humidity of air max. 80 % (not condensing)
Housing (B x D x H) mm 165 x 230 x 80
Weighing plate mm Ø 81 Ø 81 Ø 105
Weight kg (net) 1.0 1.0 1.0
Underfloor weighing - - Standard
Flush-mounted hook - - Option

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN 440 Series and is the answer not in the manual?

KERN 440 Series Specifications

General IconGeneral
Weighing plate materialstainless steel
Readability0.01g
CalibrationExternal
DisplayLCD
Reproducibility0.01 g - 1 g
Minimum part weight at piece counting (Laboratory scale)0.01g
Minimum part weight at piece counting (Precision scale)0.01g
Power supply/adapter includedincluded
Weighing unitsg, kg, oz, lb
InterfaceRS232
Unitsg, kg, oz, lb
Stabilisation time under laboratory conditions2 seconds