56 CFS/CCS-BA-e-1624
Menu settings:
Call up menu
Switch-on balance and during the selftest press .
The first menu block is displayed.
„F1 oFF“
Press repeatedly or (in CFS 50K-3 models) press
again until is displayed.
„tECH“
Acknowledge with . The request to enter the password
appears.
„Pin“
Or as standard password enter four times zero "0000" or the
stored password (input see parameter ). (Emergency
password „9999“)
Acknowledge by .
„Pin“ „----
Select bulk scales „tECH“ „rEmotE“ with .
Acknowledge with .
„tECH“ „LoCAL“
„tECH“ „rEmotE“
Select the weighing unit [kg or lb] with or (model
CFS 50K-3), where the adjustments shall be made.
The appeared [
] displays the current weighing unit.
Confirm with ; the next menu item „Cnt“ will be displayed.
tECH“ „Unit“
„Cnt“