CFS/CCS-BA-e-1624 57
(1) Adjusting the bulk scales, all models except for CFS 50K-3
1. Internal resolution
Press , the internal resolution will be shown.
Return to menu by .
Use to select the next menu item „Cap“.
„Cnt“
2. Position decimal point / capacity
CAP
Press on display „CAP“; the currently adjusted position
of the decimal point will be displayed.
Select desired setting with and acknowledge by .
The currently set capacity is displayed.
Delete with for changes, and input desired value with
the numeric keys.
Confirm input with , the balance returns to the menu.
Select the next menu point „div“ with .
„dESC“ „0.00“
„SEL“ „000030“
„CAP“
3. Readability
„div“
Press , the current setting will be displayed.
Select desired setting by and confirm with , the
balance returns to the menu.
Select the next menu point „AZt“ with
„inC“ „1“
„div“
4. Automatic zero tracking
on display change
AZt
Press , the current setting will be displayed.
Select desired setting by and confirm with , the
balance returns to the menu.
Select the next menu point „0 AUto“ with .
„AZn“ „2d“
„AZt“