KN-4006 series UV Radiation Treatment System
29
hospital in accordance with the provisions of calibration cycle,
when the calibration cycle ends, please contact the
manufacturer. Recommend that users inspect equipment and
accessories in their daily use in order to receive appropriate
treatment.
8. Equipment and accessories service life expires, please deal
with it in accordance with the related electronic products waste
handling requirements.
5.2 Equipment cleaning
Warning: Cut off the AC power. before cleaning
equipment.
1. The most commonly used hospital cleaning fluid and
non-corrosive detergents can be used to clean the equipment,
but note that many of them need to be diluted before use,
please follow the detergent manufacturer's instructions to use.
2. Avoid the using ethanol-based, amino-or acetone-based
cleaning agents.
3. Lamp housing, screen, lamp should be kept free of dust
pollution, soft lint-free cloth or sponge with the cleaning agent
can be used to wipe.
Notice: Be careful that not to pour the liquid on the
equipment,
4. Prohibiting the use of such steel wire brush or metal polishing