KN-4006 series UV Radiation Treatment System
31
Transport and storage environmental conditions:
Ambient temperature: -40 ~ 55
Atmospheric pressure: 500 ~ 1060hpa
Relative humidity: ≤ 95%
5.5 Lamp Replacement
UVA, UVB lamp was produced by Philips, the power is 9W.
The normal life of UV lamp is 1000 hours (this data is
provided by the manufacturers ), but with the increased use of
time, frequent switches or too high temperature will lead to
reduce the intensity and shorten lamp life, when the lamp flashes
or does not work, please change the light. Meanwhile, in order to
ensure the effectiveness of treatment, it is suggested to change
new lamps when the use time is up to 350 hours or an obvious
drop on intensity occurs (less than 50% of the original data),to
keep a better therapeutic effect.
Notice:
- The same type of lamp must be replaced in order to
ensure the normal equipment.
- Lamps are fragile items, in the process of demolition
and installation it is important to note the intensity is
moderate with handle care and put down gently.
- The lamp should be discarded in accordance with
special waste for processing.