KN-4006 series UV Radiation Treatment System
40
Table B-4 Narrow Spectrum UVB Dose
Skin Type
The Initial Dose
J/cm²
Increasing Dose
J/cm²
Maximum Dose
J/cm²
I
0.2
0.05
2
II
0.2
0.05
2
III
0.3
0.10
3
IV
0.3
0.10
3
V
0.4
0.15
5
VI
0.4
0.15
5
Notice: - Although the method that determine radiation
dosage by skin type is simple, but not as accurate
as MPD or MED.
- Specific drugs may increase the skin sensitivity to
light, thereby to affect the determination of skin
type.
- It is also important to learn about the patient’s
reaction after long-term sun exposure to
determine radiation dose. If two patients are the
same skin type, the slow tanned one should
adopts a lower dose.