KN-4006 series UV Radiation Treatment System
73
Recommended separation distances between portable and mobile RF communications equipment and
the EQUIPMENT or SYSTEM - for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Recommended separation distances between
portable and mobile RF communications equipment and The KN-4000 series UV Radiation Treatment System
The KN-4000 series UV Radiation Treatment System is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of The KN-4000 series UV Radiation Treatment System can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and The KN-4000 series UV Radiation Treatment System as recommended below, according to the maximum output
power of the communications equipment
Rated maximum
output of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
P
V
d ]
5.3
[
1
=
P
E
d ]
5.3
[
1
=
P
E
d ]
7
[
1
=