Introduction 1
Keysight E-Series E9300 Operating and Service Guide 27
Recommended calibration interval
Keysight Technologies recommends a one-year calibration cycle for the E-Series
E9300 power sensors.
Specifications
The specifications listed in Chapter 3, "Characteristics and Specifications", are the
performance standards or limits against which the power sensor may be tested.
These specifications are valid ONLY after proper calibration of the power meter.
Refer to the “Calibration Procedure Using Keysight E-Series Power Sensors” in
your Keysight EPM series power meter User’s Guide.
Use the plastic flange cover to protect the waveguide connector from dirt and
mechanical damage whenever it is not in use. Any burn, dents or dirt on the
flange or waveguide surface will increase the SWR and change the Cal Factor.
Refer to the power meter operating and service manual for interconnecting
instructions.