CALibration:STRing "<string>"
CALibration:STRing?
Record or read up to 40 characters of instrument calibration information.
You can store any information related to calibration, such as date or contact information.
The power supply must be unsecured before sending a calibration message.
You can read the string regardless of whether the supply is secured.
CALibration:VOLTage[:DATA] <value>
Enters an output voltage value (in volts) that you obtained by reading a DMM.
You must first select a calibration level (CAL:CURR:LEV) for the value being entered.
This command can only be used when calibration is unsecured and output is ON.
CALibration:VOLTage:LEVel MINimum|MAXimum
Initiates calibration for a specific voltage level. You must do the MINimum calibration before the MAXimum, and this
command can only be used when calibration is unsecured.
47
Keysight E36100B Series Operating and Service Guide