Test Record Forms
Test Record Form - Keysight E36102B
Test Record Form - Keysight E36103B
Test Record Form - Keysight E36104B
Test Record Form - Keysight E36105B
Test Record Form - Keysight E36106B
Test Record Form - Keysight E36102B
E36102B Report Number ____________ Date ____________
Description Lower limit Result Upper limit
Constant Voltage Tests
Voltage Programming
Zero Voltage Output (V
0
)
Maximum Voltage Output (V
max
)
-0.003 V
5.994 V
______________
______________
0.003 V
6.006 V
Voltage Readback
Zero Voltage measured over interface
Maximum Voltage measured over interface
V
0
- 0.003 V
V
max
- 0.006 V
______________
______________
V
0
+ 0.003 V
V
max
+ 0.006 V
CV Load Regulation (V
load
- V
noload
) -2.6 mV ______________ 2.6 mV
CV Line Regulation (V
lowline
- V
highline
) -1.6 mV ______________ 1.6 mV
CV Ripple and Noise
peak-to-peak
rms
______________
______________
8 mV
350 µV
Transient Response ______________ 50 μs
Constant Current Tests
Correct Programming
Zero Current Output (I
0
)
Maximum Current Output (I
max
)
-0.004 A
4.9925 A
______________
______________
0.004 A
5.0075 A
Current Readback
Zero Current measured over interface
Maximum Current measured over interface
I
0
- 0.004 A
I
max
- 0.0065 A
______________
______________
I
0
+ 0.004 A
I
max
+ 0.0065 A
78
Keysight E36100B Series Operating and Service Guide