Calibration security
The instrument has a calibration security code to prevent accidental or unauthorized calibration. When you receive
your power supply, it is secured by a default security code, as shown below. The security code is not changed by a
power cycle or *RST.
Model Default code
E36102B 036102
E36103B 036103
E36104B 036104
E36105B 036105
E36106B 036106
You can enter a security code of up to 6 digits on the front panel or up to 9 digits from the remote interface.
To reset a forgotten security code to its factory default, you can turn on the instrument after shorting
CAL SECURE jumper J3 inside the instrument as shown below.
Keysight E36100B Series Operating and Service Guide 89