Keysight Technologies E36100B Series - Voltage Subsystem; [Source:]Voltage[:Level][:Immediate][:Amplitude] <Voltage>|Minimum|Maximum|Default [Source:]Voltage[:Level][:Immediate][:Amplitude]? [Minimum | Maximum]; [Source:]Voltage[:Level][:Immediate]:Step[:Increment] <Numeric Value>|Default[Source:]Voltage [:Level][:Immediate]:Step[:Increment]? [Default]; [Source:]Voltage[:Level]:Triggered[:Amplitude] <Voltage>|Min|Max[Source:]Voltage[:Level]:Triggered [:Amplitude]? [MIN|MAX]

Keysight Technologies E36100B Series
93 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
VOLTage Subsystem
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] <voltage>|MINimum|MAXimum|DEFault
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]? [MINimum | MAXimum]
Sets the output voltage in volts.The query returns a number of the form +#.########E+##.
[SOURce:]VOLTage[:LEVel][:IMMediate]:STEP[:INCRement] <numeric value>|DEFault
[SOURce:]VOLTage[:LEVel][:IMMediate]:STEP[:INCRement]? [DEFault]
Sets the step size for voltage programming with the VOLTage UP and VOLTage DOWN commands. The query
returns a number of the form +#.########E+##.
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude] <voltage>|MIN|MAX
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]? [MIN|MAX]
Sets the triggered voltage level in amps. The query returns a number of the form +#.########E+##.
[SOURce:]VOLTage:PROTection:CLEar
Clears an overvoltage protection event.
[SOURce:]VOLTage:PROTection:STATe ON|1|OFF|0
[SOURce:]VOLTage:PROTection:STATe?
Enables or disables overvoltage protection, which causes the instrument to go into a protected state when the
power supply goes above the protection level specified by the VOLTage:PROTection[:LEVel]. The query returns 1
(ON)or 0 (OFF).
[SOURce:]VOLTage:PROTection:TRIPped?
Indicates whether an overvoltage protection occurred (1) or not (0). This is reset to 0 by
VOLTage:PROTection:CLEar.
[SOURce:]VOLTage:PROTection[:LEVel] <voltage>|MINimum | MAXimum[SOURce:]VOLTage:PROTec-
tion[:LEVel]? [MINimum|MAXimum]
Sets the level at which overvoltage protection trips, in volts. The query returns a number of the form
+#.########E+##.