Keysight Technologies E36100B Series - Current Subsystem; [Source:]Current[:Level][:Immediate][:Amplitude] <Current> | Minimum | Maximum | up | down [Source:]Current[:Level][:Immediate][:Amplitude]? [Minimum | Maximum]; [Source:]Current[:Level][:Immediate]:Step[:Increment] <Current> | Default[Source:]Current[:Level] [:Immediate]:Step[:Increment]? [Default]; [Source:]Current[:Level]:Triggered[:Amplitude] <Current> | Minimum | Maximum[Source:]Current [:Level]:Triggered[:Amplitude]? [Minimum | Maximum]

Keysight Technologies E36100B Series
93 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CURRent Subsystem
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] <current> | MINimum | MAXimum |UP | DOWN
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]? [MINimum | MAXimum]
Programs the immediate current level of the instrument's output. See Programming Ranges for details.
This command changes the output immediately. The UP and DOWN parameters increase or decrease the
immediate current by the amount specified by CURRent:STEP. A setting that exceeds the maximum or minimum
rated current will cause error -222 (Data out of range).
[SOURce:]CURRent[:LEVel][:IMMediate]:STEP[:INCRement] <current> | DEFault
[SOURce:]CURRent[:LEVel][:IMMediate]:STEP[:INCRement]? [DEFault]
Sets the step size for current programming with the CURRent UP and CURRent DOWN commands. The query
returns a number of the form +#.########E+##.
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude] <current> | MINimum | MAXimum
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]? [MINimum | MAXimum]
Sets the triggered current level in amps. The query returns a number of the form +#.########E+##.
[SOURce:]CURRent:PROTection:CLEar
Clears an overcurrent protection event.
[SOURce:]CURRent:PROTection:DELay[:TIME] <time> | MINimum |MAXimum
[SOURce:]CURRent:PROTection:DELay[:TIME]? [MINimum |MAXimum]
Sets the time (in milliseconds) that the overcurrent protection is temporarily disabled after a current level change.
The query returns a number of the form +#.########E+##.
[SOURce:]CURRent:PROTection:STATe ON|1|OFF|0
[SOURce:]CURRent:PROTection:STATe?
Enables or disables overcurrent protection, which causes the instrument to go into a protected state when the
power supply status is in constant current mode. The query returns 1 (ON)or 0 (OFF).
[SOURce:]CURRent:PROTection:TRIPped?
Indicates whether an overcurrent protection occurred (1) or not (0). This is reset to 0 by
CURRent:PROTection:CLEar.
Keysight E36100B Series Operating and Service Guide 48