Remote Interface Reference 3
Keysight E364xA User’s and Service Guide 113
To ensure synchronization when the bus source is selected, send the *WAI
(wait) command. When the *WAI command is executed, the power supply
waits for all pending operations to complete before executing any additional
commands. For example, the following command string guarantees that the
first trigger is accepted and is executed before the second trigger is
recognized.
You can use the *OPC? (operation complete) command or the *OPC
(operation
complete) command to signal when the operation is complete. The *OPC?
command returns “1” to the output buffer when the operation is complete. The
*OPC command sets the “OPC” bit (bit 0) in the Standard Event register when
the operation is complete.
Immediate triggering
To select the immediate trigger source, send the following command.
–When the IMMediate is selected as a trigger source, an INITiate command
immediately transfers the VOLT:TRIG or
CURR:TRIG value to VOLT or CURR
value. Any delay is ignored.
TRIG:SOUR BUS;*TRG;*WAI;*TRG;*WAI
TRIG:SOUR IMM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E3644A and is the answer not in the manual?

Keysight Technologies E3644A Specifications

General IconGeneral
BrandKeysight Technologies
ModelE3644A
CategoryPower Supply
LanguageEnglish

Related product manuals