8 Service and Maintenance
220 E364xA User’s and Service Guide
Performance Test Record
Table 8-6 CV performance test record
Test description Model Actual result Specification
Upper limit Lower limit
CV programming accuracy at
0 V (DVM reading)
all +0.0100 V -0.0100 V
CV readback accuracy at 0 V all DVM +0.0050 V DVM –0.0050 V
CV programming accuracy at
Full Scale (DVM reading)
E3640A/E3642A/E3644A +20.0200 V +19.9800 V
E3641A/E3643A/E3645A +60.0400 V +59.9600 V
CV readback accuracy at
Full Scale
E3640A/E3642A/E3644A DVM +0.0150 V DVM -0.0150 V
E3641A/E3643A/E3645A DVM +0.0350 V DVM -0.0350 V
CV load effect (load regulation) E3640A/E3642A/E3644A Maximum change: <5 mV
E3641A/E3643A/E3645A Maximum change: <9 mV
CV source effect (line
regulation)
E3640A/E3642A/E3644A Maximum change: <5 mV
E3641A/E3643A/E3645A Maximum change: <9 mV
CV PARD (normal mode) E3640A/E3642A/E3644A <5 mVpp / 0.5 mVrms
E3641A/E3643A/E3645A <8 mVpp / 1 mVrms
Load transient response time all <50 µs