3 Remote Interface Reference
114 Keysight E364xA User’s and Service Guide
Triggering commands
INITiate
This command causes the trigger system to initiate. This command completes
one full trigger cycle when the trigger source is an immediate and initiates the
trigger subsystem when the trigger source is bus.
TRIGger:DELay {<seconds>| MINimum|MAXimum}
This command sets the time delay between the detection of an event on the
specified trigger source and the start of any corresponding trigger action on
the power supply output. Select from 0 to 3600 seconds. MIN
= 0 seconds. MAX
= 3600 seconds. At *RST
, this value is set to 0 seconds.
TRIGger:DELay?
Query the trigger delay.
TRIGger:SOURce {BUS|IMMediate}
Selects the source from which the power supply will accept a trigger. The
power supply will accept a bus (software) trigger or an internal immediate
trigger. At *RST, the bus trigger source is selected.
TRIGger:SOURce?
Query the present trigger source. Returns “BUS” or “IMM”.
*TRG
Generate a trigger to the trigger subsystem that has selected a bus (software)
trigger as its source (TRIG:SOUR BUS). The command has the same effect as
the Group Execute Trigger (GET) command. For RS-232 operation, make sure
the power supply is in the remote interface mode by sending the SYST:REM
command first.