Remote Interface Reference 3
Keysight E364xA User’s and Service Guide 135
*ESE<enable value>
Enable bits in the Standard Event enable register. The selected bits are then
reported to the Status Byte.
*ESE?
Query the Standard Event enable register. The power supply returns a decimal
value which corresponds to the binary-weighted sum of all bits in the register.
*ESR?
Query the Standard event register. The power supply returns a decimal value
which corresponds to the binary-weighted sum of all bits in the register.
*OPC
Set the “Operation Complete” bit (bit 0) of the Standard Event register after the
command is executed.
*OPC?
Return “1” to the output buffer after the command is executed.
*PSC { 0|1 }
This power-on status clear command clears the Status Byte and the Standard
Event register enable masks when power is turned on (*PSC 1). When *PSC 0
is in effect, the Status Byte and Standard Event register enable masks are not
cleared when power is turned on.
*PSC?
Query the power-on status clear setting. The returned parameter is “0” (*PSC
0) or “1” (*PSC 1).
*SRE <enable value>
This command enables bits in the Status Byte enable register.
*SRE?
Query the Status Byte Enable register. The power supply returns a decimal
value which corresponds to the binary-weighted sum of all bits set in the
enable register.