118 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Optimizing Performance for All Models
Using External Leveling (N5173B/83B Only)
4. Determine the output amplitude range, see “Determining the Signal Generator’s Amplitude
Range” on page 119
5. Set the displayed power value, see “Adjusting the Signal Generator Display’s Amplitude
Value” on page 120
Equipment Setup
Set up the equipment as shown in Figure 5-22 on page 118. Place the external detector (detector
and coupler/power splitter) as close as possible to the DUT.
Recommended Equipment
Keysight 8474E negative detector
Keysight 87301D directional coupler
cables and adapters, as required
Figure 5-22 Typical External Leveling Setup using a Directional Coupler
Configuring the Carrier
1. Press Preset.
2. Set the carrier frequency.
3. Set the power level to 0 dBm:
If the signal generator has Options 1E1 and 520, set the output attenuator to zero dBm:
a. Press AMPTD > Atten/ALC Control > Atten Hold Off On to On.
b. Press Set Atten > 0 > dB.
c. Press Set ALC Level > 0 > dBm.
Selecting External Leveling
Press AMPTD > Leveling Control > Leveling Mode > Pwr Meter Cont.
Leveled Signal
Signal Generator
Coupler
Amplifier
Negative Detector
RF OUTPUT
ALC INPUT

Table of Contents

Other manuals for Keysight Technologies N5173B EXG

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N5173B EXG and is the answer not in the manual?

Keysight Technologies N5173B EXG Specifications

General IconGeneral
BrandKeysight Technologies
ModelN5173B EXG
CategoryPortable Generator
LanguageEnglish

Related product manuals