Keysight E5052B Signal Source Analyzer 13
Installation
Environmental Requirements
Environmental Requirements
Set up the E5052B where the following environmental requirements are met.
Operating Environments
Ensure that the operating environment meets the following requirements.
Temperature 10ºC to 40ºC
Temperature range at the
error-correction
23ºC 5ºC (< 1ºC deviation from the temperature
when performing the error-correction)
Humidity 20% to 80% at wet bulb temperature < +29ºC
(non-condensation)
Altitude 0 to 2,000 m (0 to 6,561 feet)
Vibration 0.21 G maximum, 5 Hz to 500 Hz
Above environmental requirements are NOT for the specifications and
measurement accuracy of the analyzer, but for the operating environment
of the analyzer.