14 Keysight E5052B Signal Source Analyzer
Installation
Environmental Requirements
2-
Ventilation Requirements
To ensure safety requirements, the specifications and measurement accuracy
of the analyzer, you must keep environmental temperature within the specified
range by providing appropriate cooling clearance around the analyzer or, for
the rackmount type, by forcefully air-cooling inside the rack housing. For more
information on environmental temperature to satisfy the specifications and
measurement accuracy of the analyzer, see the specification in the E5052B
Online Help.
When the environmental temperature around the analyzer is kept within the
temperature range of the operating environment specification (See Section
“Operating Environments” on page 13), the analyzer conforms to the
requirements of the safety standard. Furthermore, under that temperature
requirement, the analyzer still conforms to the requirements of the safety
standard even when placing the analyzer with cooling clearance as follows:
Figure 2-1 Ventilation space at the installation
Requirements
Back ≥ 180 mm
Sides ≥ 60 mm (both right and left)
When measuring place the E5052B in a proper position as show in Figure
2-1.