Getting Started Guide 15
Safety & Environmental Information
Instrument Location and Rack Mounting Requirements
Instrument Location and Rack Mounting Requirements
Locating the Test Platform
Make sure that the left-side panel fan inlet and right-side panel exhaust vent
areas are not obstructed. The minimal required clearance is 2.75 inches (7 cm).
Table Top Ambient Temperature
Rack Mounting: Hardware and Temperature
If you choose to locate your test platform in a rack, follow the guidelines
provided in this section.
Based on the type of equipment rack you have, you must determine what rack
rails you need. If you are using a Keysight System Test Rack, you can find
information on what to order by referring to the Rack Mounting Flange Kit
(Option E7515B-1CM) Installation Note.
Install the instrument so that the detachable power cord is readily identifiable
and is easily reached by the operator. The detachable power cord is the
instrument disconnecting device. It disconnects the mains circuits from the
mains supply before other parts of the instrument. The front-panel switch is only
a standby switch and does not act as a LINE switch. If needed, an externally
installed switch or circuit breaker (which is readily identifiable and is easily
reached by the operator) may be used as a disconnecting device.
Do not exceed an ambient temperature of 40° C when operating the instrument
on a table top.
When mounting instrument in a rack, do not exceed the level of:
—Internal rack air temperature of 40°C