162 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
15 Measurements
Base
The Base of a waveform is the mode (most common value) of the lower part of the
waveform, or if the mode is not well defined, the base is the same as Minimum.
The Y cursor shows the value being measured.
Overshoot
Overshoot is distortion that follows a major edge transition expressed as a
percentage of Amplitude. The X cursors show which edge is being measured (edge
closest to the trigger reference point).
Figure 33 Isolating area for Top measurement
Rising edge overshοot =
local Maximum D Top
Amplitude
× 100
Falling edge overshοot =
Base D local Minimum
Amplitude
× 100