Vertical Controls 3
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 51
 
When bandwidth limit is on, the maximum bandwidth for the channel is 
approximately 20 MHz. For waveforms with frequencies below this, turning 
bandwidth limit on removes unwanted high frequency noise from the waveform. 
The bandwidth limit also limits the trigger signal path of any channel that has 
BW Limit turned on.
To change the vertical scale knob's coarse/fine adjustment setting
1 Push the channel's vertical scale knob (or press the channel key and then the 
Fine softkey in the Channel menu) to toggle between fine and coarse 
adjustment of the vertical scale.
When Fine adjustment is selected, you can change the channel's vertical 
sensitivity in smaller increments. The channel sensitivity remains fully calibrated 
when Fine is on.
The vertical scale value is displayed in the status line at the top of the display.
When Fine is turned off, turning the volts/division knob changes the channel 
sensitivity in a 1-2-5 step sequence.
To invert a waveform
1 Press the desired channel key.
2 In the Channel menu, press the Invert softkey to invert the selected channel.
When Invert is selected, the voltage values of the displayed waveform are inverted.
Invert affects how a channel is displayed. However, when using basic triggers, the 
oscilloscope attempts to maintain the same trigger point by changing trigger 
settings.
Inverting a channel also changes the result of any math function selected in the 
waveform Math menu or any measurement.
Setting Analog Channel Probe Options
1 Press the probe's associated channel key.