Service Guide N5227-90001 2-5
PNA Series Microwave Network Analyzers      General Product Information
N5227A Required Service Test Equipment
Required Service Test Equipment
Equipment
a
a. Unless specified otherwise, equipment listed is required for all analyzer models.
Critical Specifications 
Recommended 
Model or Part 
Number
Alternate Model 
or Part Number
Use
b
b. P = Performance tests, A = Adjustments, T = Troubleshooting, V = System verification
Test Instruments and Software
Compression
test set
None specified
U3070AK01
None P
Dynamic accuracy
test set
None specified
U3020AD01
None P
Tes t s oftwa re
c
c. The recommended model or part number for all equipment listed with a “P” in the Use column is required for proper operation of 
this test software.
N/A N7840A None
P
Frequency counter
Freq: 10 MHz to 20 GHz 
Accuracy : 0.5 ppm
53151A
Opt 001
None P, A,T
Signal generator CW Freq: 1.185 GHz
N5181A, 
E8257D, 
Option 520
P, A, T 
Power meter Accuracy: ±0.0068 dB N1913A/14A
E4418A/B, 
E4419A/B
d
d. If an accurate measurement of the dynamic accuracy specification is not required, the E4418A or E4419A can be used.
P, A,T
Power sensor Freq: 10 MHz to 4.2 GHz 
Range: –30 to +20 dBm
N8482A
8482A
P, A,T
Power sensor
Freq: 50 MHz to 40 or 50 GHz
Range: –30 to +20 dBm
N8487A
8487A
P, A,T
Power sensor
e
e. This sensor and adapter must be calibrated together by the Keysight factory. The data supplied is only valid as long as the sensor 
and adapter remain connected.
Freq: 50 MHz to 67 GHz
Range: –30 to +20 dBm
V8486A
None
P, A,T
Adapter
e
(WR-15 to 1.85 mm)
Return Loss: 
 22 dB
V281A
None
P, A,T
Spectrum analyzer
Min Freq: 1 MHz
Max Freq: > 4 GHz
Resolution BW: 300 Hz
E444xA PSA series, 
N90xxA signal 
analyzer family
856xE A,T
Digital voltmeter Resolution: 10 mV Any Any T
Printer N/A
Any printer with Microsoft Windows XP 
or Windows 7 driver
--
Mouse N/A Any Any --
Keyboard N/A Any Any --