EasyManua.ls Logo

Keysight N5227A - Enabling or Removing Options

Keysight N5227A
326 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8-14 Service Guide N5227-90001
General Purpose Maintenance Procedures PNA Series Microwave Network Analyzers
Option Enable Utility N5227A
Enabling or Removing Options
There are two types of options:
Hardware: Hardware options involve adding additional hardware to the analyzer. After the proper
hardware has been installed in the analyzer, the option can be enabled using the option enable utility.
It is necessary to re-enable all installed hardware options if the test set motherboard is replaced.
Software: Software options add features or functionality to the analyzer without the need for additional
hardware. These options are enabled using the option enable utility and require a special license key.
NOTE Some applications require a license key that is provided by Keysight. If you do not have the
required license key, contact Keysight for assistance. Refer to “Contacting Keysight” on
page 2-6.
It is necessary to backup all installed software options if the frequency reference board is replaced.
To enable or remove an option:
1. Start the option enable utility. Refer to Accessing the Option Enable Utility” on page 8-13.
2. Click the arrow in the
Select Desired Option box. A list of available options, similar to the list below, will
appear.
020 - Add IF inputs
021 - Pulse Modulator on Source 1
022 - Pulse Modulator on Source 2
025 - 4-Receiver Pulse Generator
086 - Gain Compression Application
087- IMD Measurements
219 - Src/Rcvr Atten & Bias Tees 2-Port
419 - Src/Rcvr Atten & Bias Tees 4-Port
224 - 2nd Src w/Combiner & Switches
423 - Combiner & Switches
??? - Enter Unlisted Option
3. Click on the option that you wish to either enable or remove, and then click
Enable or Remove, whichever
is appropriate.
4. If the desired option is not available in the list, select
??? - Enter Unlisted Option. A dialog box appears that
will allow you to enter the option number. Enter the option number and follow the instructions on the
display.