3-46 Service Guide N5227-90001
Tests and Adjustments PNA Series Microwave Network Analyzers
Adjustments N5227A
Figure 3-17 Equipment Setup 3 for the Receiver Adjustment
Receiver Characterization
This characterizes the receivers in your analyzer.
Procedure
1. Press UTILITY , then , then , then click Receiver Characterization.
2. Follow the instructions and prompts as they are displayed.
EE Default Adjustment
This sets the EEPROM data to their default values.
Procedure
1. Press UTILITY , then , then , then click EE Default Adjustment.
2. Follow the instructions and prompts as they are displayed.