Service Guide N5227-90001 3-45
PNA Series Microwave Network Analyzers Tests and Adjustments
N5227A Adjustments
Procedure
1. Connect the equipment as shown in Figure 3-15. Connect a GPIB cable between the network analyzer
and the power meter.
Figure 3-15 Equipment Setup 1 for the Receiver Adjustment
2. Press UTILITY , then , then , then click Receiver Adjustment.
3. Ensure the GPIB settings are correct.
4. Follow the instructions and prompts as they are displayed.
Figure 3-16 Equipment Setup 2 for the Receiver Adjustment