7-16 Service Guide N5227-90001
Repair and Replacement Procedures PNA Series Microwave Network Analyzers
Removing and Replacing the A20 Power Supply Assembly N5227A
Removing and Replacing the A20 Power Supply Assembly
Tools Required
• T-10 TORX driver (torque to 9 in-lb or 1.02 N.m)
• T-20 TORX driver (torque to 21 in-lb or 2.38 N.m)
• ESD grounding wrist strap
Removal Procedure
Refer to Figure 7-3 for this procedure.
1. Disconnect the power cord.
2. Remove the outer and inner covers. Refer to “Removing the Covers” on page 7-6.
3. Remove the three flat head screws (item
①
) from the power supply bracket.
4. Remove the seven pan head screws (item
➁) from the power supply rear panel.
5. Slide the A20 power supply assembly out the rear of the analyzer.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Perform the post-repair adjustments, verifications, and performance tests that pertain to this removal
procedure. Refer to Table 7-2 on page 7-54.