Grouped-Channel View
Channels 2 through 4 are connected in parallel and have been configured or grouped to act as a
single, higher-power channel
Refer to Output Groups for more
information
Grouped channels are addressed using the channel number of the lowest channel in the group
Double-Wide View
Channel 2 is a double-wide power module that occupies two channel locations in the mainframe
Operating status
indicators
OFF = the output is off
CV = the output is in constant voltage mode
CC = the output is in constant current mode
CP+ = the output is limited (or disabled) by the positive power limit
CP– = the output is limited (or disabled) by the negative power limit
VL+/– = the output is in positive or negative voltage limitmode
CL+/– = the output is in positive or negative current limitmode
OV = the output is disabled by the over-voltage protection
OV– = the output is disabled by the negative over-voltage protection
OC = the output is disabled by the over-current protection
OT = the over-temperature protection has tripped
PF= the output is disabled by a power-fail condition
OSC = the output is disabled by the oscillation protection
INH = the output is disabled by an external inhibit signal
UNR = the output is unregulated
PROT = the output is disabled by a condition from a coupled channel
SH = the load input terminals are shorted (N679xA)
UVI = an under-voltage inhibit condition occurred (N679xA)
Interface status
indicators
Err = an error has occurred (press Error key to display error message)
Lan = the LAN is connected and has been configured
IO = there is activity on one of the remote interfaces
Keysight N6700C Operating and Service Guide 21
1 Quick Reference