Command Quick Reference
Some [optional] commands have been included for clarity. All settings commands have a
corresponding query. See the Syntax Conventions for SCPI.
ABORt (ACQuire and ELOG commands only on N676xA, N678xA SMU, Option 054)
îîîîîî:ACQuire (@chanlist) Cancels any triggered measurements.
îîîîîî:ELOG (@chanlist) Stops external data logging.
îîîîîî:TRANsient (@chanlist) Cancels any triggered actions.
CALibrate
îîîîîî:COUNt? Returns the number of times the unit has been calibrated.
îîîîîî:CURRent
[:LEVel] <value>, (@channel) Calibrates the current programming.
îîîîîîîîîîîî:LIMit
îîîîîîîîîîîîîîîîîî:NEGative <value>, (@channel) Calibrates the negative current limit. (N678xA SMU, N6783A-BAT)
îîîîîîîîîîîîîîîîîî:POSitive <value>, (@channel) Calibrates the positive current limit. (N678xA SMU, N6783A, N679xA)
îîîîîîîîîîîî:MEASure <value>, (@channel) Calibrates the current measurement.
îîîîîîîîîîîî:PEAK (@channel) Calibrates the peak current limit. N675xA, N676xA)
îîîîîî:DATA <value> Enters the calibration value read by the external meter.
îîîîîî:DATE <"date">, (@channel) Stores the calibration date in nonvolatile memory.
îîîîîî:DPRog (@channel) Calibrates the current downprogrammer.
îîîîîî:LEVel P1|P2|P3 Advances to the next level in the calibration.
îîîîîî:PASSword <value> Sets a numeric password to prevent unauthorized calibration.
îîîîîî:RESistance 20|6, (@channel) Calibrates the output resistance. (N6781A, N6785A)
îîîîîî:SAVE Saves the calibration constants in non-volatile memory.
îîîîîî:STATe 0|OFF|1|ON Enables/disables calibration mode.
îîîîîî:VOLTage
[:LEVel] <value>, (@channel) Calibrates the voltage programming.
îîîîîîîîîîîî:CMRR, (@channel) Calibrates the voltage common mode rejection ratio. (N675xA, N676xA)
îîîîîîîîîîîî:LIMit
îîîîîîîîîîîîîîîîîî:POSitive <value>, (@channel) Calibrates the positive current limit. (N678xA SMU)
îîîîîîîîîîîî:MEASure <value>, (@channel) Calibrates the voltage measurement.
îîîîîîîîîîîîîîîîîî:AUXiliary, (@channel) Calibrates the auxiliary voltage measurement. (N6781A, N6785A)
DISPlay
[:WINDow]
[:STATe] 0|OFF|1|ON Turns the front panel display on or off.
îîîîîîîîîîîî:CHANnel <channel> Selects the channel in 1-channel view.
îîîîîîîîîîîî:VIEW METER1|METER4|METER_VIP Selects 1-channel, 4-channel, or 1 channel with power view.
Keysight N6700C Operating and Service Guide 27
1 Quick Reference