Configuring User-Defined Arbs
User-Defined Arbs can only be programed using SCPI commands. Refer to User-
Defined commands .
Resistance and Powerntypes apply to . The parameters apply to the
selected Arb type (Voltage, Current, Power, or Resistance).
Front Panel Menu Reference SCPI Command
Not available To set up the Arb type (current, voltage, power, or resistance) and shape, use:
ARB:FUNC:TYPE VOLT|CURR|POW|RES,(@1)
ARB:FUNC:SHAP UDEF,(@1)
To specify the voltage value of five steps:
ARB:VOLT:UDEF:LEV 1,2,3,4,5,(@1)
To specify the time that the output stays at the voltage steps:
ARB:VOLT:UDEF:DWEL 1,2,3,2,1,(@1)
To generatean external trigger signal at the start of a voltage step
(the trigger is generated atthe start of step #3):
ARB:VOLT:UDEF:BOST 0,0,1,0,0,(@1)
Configuring Steps Common to all Arbs
Step 4. Specify what happens when the Arb completes.
Select whether the output returns to the DC value that was in effect before the Arb started, or
whether the output should remain at the last Arb value.
Front Panel Menu Reference SCPI Command
Select Transient\Arb\Terminate.
Specify Return to start setting or Stop atlast Arb setting.
Then press Select.
To specify a setting when the Arb ends, use:
ARB:TERM:LAST OFF,(@1)
Step 5. Specify the repeat count of the Arb.
Specify if the Arb should repeat, either continuously (INFinity), or only for a specified repeat count. A
repeat count of 1 runs the Arb once.
Front Panel Menu Reference SCPI Command
Select Transient\Arb\Repeat.
Specify a repeatcount, or run the Arb continuously. Then
press Select.
To specify an Arb count, use:
ARB:COUN 1,(@1)
4 Operating the Power System
128 Keysight N6700C Operating and Service Guide