Calibration Procedure – Models N6791A and N6792A
Enter Calibration Mode
Front Panel Menu Reference SCPI Command
Select System\Admin\Login.
Enter your password. Then press Select.
CAL:STAT ON <password>
Voltage Calibration
High Range Measurement
Step 1. • Connect an external power supply to the load module’s input terminals. Set the power supply
to 60 V with a current limit of 0.5 A. Connect a DVM capable of handling 60 V in parallel with the unit’s
input (see Cal Setup G). Turn on the output of the power supply.
Step 2. Select the full-scale voltage range.
Front Panel Menu Reference SCPI Command
Select System\Admin\Cal\Volt\Vprog.
Select the High range. Then press Next.
CAL:VOLT 60, (@1)
Step 3. Select the first voltage calibration point and enter the data. Wait a minimum of 1 minute for the
current to stabilize before entering the data.
Front Panel Menu Reference SCPI Command
Display shows: “Enter P1 measured data”Enter the data from the external
DMM.Press Select when done.
CAL:LEV P1
*OPC?
CAL:DATA <data>
Step 4. Select the second voltage calibration point and enter the data. Wait a minimum of 1 minute for
the current to stabilize before entering the data.
Front Panel Menu Reference SCPI Command
Display shows: “Enter P2 measured data”Enter the data from the external
DMM.Press Select when done.
Press Next to finish calibration.
CAL:LEV P2
*OPC?
CAL:DATA <data>
Keysight N6700C Operating and Service Guide 373
6 Verification and Calibration