Non-Volatile Settings
The following table shows the as-shipped settings of the non-volatile parameters. These are not
affected by power cycling or *RST.
The power-on/reset state may differ from that shown below if you have enabled power-
on state recall mode from the States menu (see Instrument State Storage).
Factory as-shipped settings
Admin/Calibration password 0 (zero)
Calibraton date March 6, 2016
Channel grouping No groups
Digital port function (all pins) Digital In
Digital port polarity (all pins) Positive
Firmware update password protected Disabled
Front panel lockout Disabled
Front panel meter view 1-channel
GPIB address 5
GPIB interface Enabled
Key clicks Enabled
LANinterface Enabled
On/Off key affects all channels Disabled
Output inhibitmode Off
Saved states *RST command
Screen saver Enabled
Screen saver delay 60 minutes
USB interface Enabled
Wake on I/O Enabled
Web server Enabled
Keysight N6700C Operating and Service Guide 317
5 SCPI Programming Reference