214                                                                                                                     Keysight N9927-90001 User’s Guide
RTSA (Real-Time Spectrum Analyzer) Mode—Option 350 (CPU2 Only)
RTSA Mode Settings
— Video A sweep is initiated when the amplitude of an incoming signal 
crosses the settable trigger level. The comparison is done in the time 
domain, before conversion to frequency domain. Refer to “Trigger Slope”, 
“Trigger Delay”, “Trigger Level” on page 215, “Auto Trigger Time” on 
page 215, and to “Trigger Hold-Off” on page 216.
— RF Burst A sweep is initiated when the signal at the third IF (analog) stage 
crosses the settable Trigger Level. This selection reacts to a wider 
bandwidth of the incoming signal than Video Trigger, which could be 
beneficial. However, RF Burst is not as sensitive to triggering on low-level 
signals. Refer to “Trigger Slope”, “Trigger Delay”, “Trigger Level” on 
page 215, “Auto Trigger Time” on page 215, and to “Trigger Hold-Off” on 
page 216.
— Periodic
 Trigger repeats at the period rate entered. Periodic trigger’s phase 
can be altered by the Offset value.
Then choose from the following:
— Period
 Enter a value for the trigger to repeat (100 ns to 400 ms). 
— Offset
 Enter a offset value to offset the phase of the periodic trigger 
(–100 ms to 100 ms).
— Reset Offset Display
 Resets Offset to default 0.000000 s.
Trigger Slope          
Trigger Slope determines which edge of an External, Video, or RF Burst trigger 
signal initiates a sweep. 
—Press Sweep 3
—Then Trigger Settings
—Then Trig Slope
— Then choose from the following:
— Pos
 Sweep is triggered by the rising (positive) edge of signal.
— Neg
 Sweep is triggered by the falling (negative) edge of signal.
Trigger Delay 
After a valid trigger signal is received, the sweep begins after the specified 
Trigger Delay time.
—Press Sweep 3
If your trigger level is set too high for the current measurement gain setup, 
you will see an error message, similar to this one, displayed: "Video 
TrigLevel was HW Limited, it is x.xx dB above current capability". Adjust 
your trigger level settings until the message is no longer displayed.