9 Strip Chart Measurement
The Strip Chart measurement is an EMI measurement for the X-
Series platform that can be used for troubleshooting in CISPR pre-
compliance and compliance test. The Strip Chart measurement is
capable of measuring a signal in real time with three EMI detectors
simultaneously. The Strip Chart measurement displays the real time
measurement results in meters in a strip chart. The update rate for
the Strip Chart measurement in both views is:
Minimum 10 times per second
Maximum 30 times per second
This topic contains the following sections:
"Measurement Commands for Strip Chart" on page 606
"Remote Command Results " on page 607
INITiate:SCHart
CONFigure:SCHart
CONFigure?
FETCh:SCHart [n]?
MEASure:SCHart [n]?
READ:SCHart [n]?