EasyManua.ls Logo

Keysight U1272A - Page 160

Keysight U1272A
188 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 Characteristics and Specifications
Electrical Specifications
160 U1271A/U1272A User’s Guide
Diode
3 V
[3]
0.0001 V 0.5% + 5 0.5% + 5
Approx.
1 mA to
2 mA
--
Auto
[4]
0.0001 V - 0.5% + 5
Approx.
0.1 mA to
0.3 mA
--
Notes for diode specifications:
1 Overload protection: 1000 Vrms for short circuits with <0.3 A current.
2 Built-in buzzer beeps continuously when the voltage measured is less than 50 mV and beeps once for forward-biased
diode or semiconductor junctions measured between 0.3 V and 0.8 V (0.3 V reading 0.8 V).
3 Open voltage for diode: <+3.3 V DC
4 Open voltage for Auto-diode: <+2.5 V DC and >–1.0 V DC
Current
300 μA
[1]
0.01 μA 0.2% + 5 0.2% + 5 - <0.04 V -
3000 μA
[1]
0.1 μA 0.2% + 5 0.2% + 5 - <0.4 V -
30 mA
[1]
0.001 mA 0.2% + 5 0.2% + 5 - <0.08 V -
300 mA
[1][3]
0.01 mA 0.2% + 5 0.2% + 5 - <1.00 V -
3 A
[2]
0.0001 A 0.3% + 10 0.3% + 10 - <0.1 V -
10 A
[2][4]
0.001 A 0.3% + 10 0.3% + 10 - <0.3 V -
Notes for DC current specifications:
1 Overload protection for 300 μA to 300 mA range: 0.44 A/1000 V; 10 × 35 mm fast-acting fuse
2 Overload protection for 3 A to 10 A range: 11 A/1000 V; 10 × 38 mm fast-acting fuse
3 Specification for 300 mA range: 440 mA continuous.
4 Specification for 10 A range: 10 A continuous. Add 0.3% to the specified accuracy when measuring signals >10 A to 20 A
for 30 seconds maximum. After measuring currents >10 A, cool down the multimeter for twice the duration of the
measured time before proceeding with low current measurements.
Table 5-1 DC specifications (continued)
Function Range Resolution
Accuracy Test current
Burden
voltage
Input
impedance
U1271A U1272A (where applicable)

Table of Contents

Other manuals for Keysight U1272A