The alarm mode “Transmitter state” is selected.
➢ Go to the “Configuration” line then press OK.
➢ Press OK on the “Condition selection” line.
The transmitter displays the list of alarms conditions :
• Ambient temperature too high*
• Ambient temperature too low*
• Memory error
• Board error*
• Probe 1 error
• Probe 2 error
• Pressure value too high*
➢ Go to the alarm condition with Up and Down arrows then press OK to select the condition.
The alarm condition is ticked.
Use the right arrow to go to the 2
nd
page of the conditions and use the Left arrow to back to the 1
st
page.
➢ Press Esc key when the required conditions are selected.
The transmitter displays a message asking to save.
➢ Go to YES then press OK.
➢ Then go to the “Audible alarm” line then press OK.
➢ Select “YES” to activate the audible alarm or “NO” not to activate it then press OK.
*Only available with the pressure board
The alarm must be completely set to be active (for example, do not keep “---” in front of a threshold, a delay,...).
3.5.5 Configure the relays (optional)
➢ Press OK.
➢ Enter the security code.
➢ Go to the “Relays” line then press OK.
The transmitter displays the 4 relays : “Relay 1”, “Relay 2”, “Relay 3” or “Relay 4”.
➢ Select the relay to configure then press OK.
➢ Press OK on “Selection” line.
➢ Select “ON”, “OFF” or “Alarm 1, 2, 3 or 4” with Up and Down arrows then press OK :
• “ON” is selected : the relay is connected. It also allows to perform a test of the relay.
• “OFF” is selected : the relay is deactivated.
• “Alarm 1, 2, 3 or 4” is selected : the activation of the relay is assigned to an alarm previously configured.
➢ Go to “Security” line then press OK.
The outputs relays are by default in negative security : the relay is energized during an alarm condition. Via the
keyboard, it is possible to switch the relays in positive security : the relay is de-energized during an alarm
condition or a power cut.
➢ Select “Positive” or “Negative” with Up and Down arrows then press OK.
To activate the relays according to the alarms, the alarms must be completely configured.
17